歡迎來到匠仁醫(yī)療設備有限公司網(wǎng)站!

更多 > 新聞資訊
公司新聞
行業(yè)知識
檢測答疑
圖標 首頁 >> 新聞資訊 > 公司新聞

公司新聞

硬管內窺鏡使用注意事項

來源:http://www.tukouzhao.cn/ 發(fā)布時間:2020-04-28 瀏覽量:0

硬管內窺鏡在手術過程中受到損壞的現(xiàn)象并不多,雖然會與人體的組織如肌肉、黏膜、骨骼等有接觸和磕碰,但是這些磕碰是輕微的,不會造成窺鏡的損壞,因為它只是起觀察的作用,不是其他器械的受力點。但是在使用其他器械時,尤其是咬合力較大的鉗、剪類器械應注意鏡管的前端不要伸進器械的咬合區(qū)內,以免誤傷鏡管。在使用這類器械時,有時醫(yī)生為了看清楚咬合區(qū)的組織,把窺鏡伸得很靠近組織,器械咬合時窺鏡沒有退回,誤傷了窺鏡。手術時如注意讓器械的咬合口全部都在窺鏡的觀察范圍內就可以避免此類事故發(fā)生。
Hard tube endoscope is not damaged in the process of operation. Although it will contact and bump with human tissues such as muscles, mucous membranes, bones, etc., these bumps are slight and will not cause damage to the endoscope, because it only plays the role of observation, not the force point of other instruments. However, when using other instruments, especially the forceps and scissors with large bite force, it should be noted that the front end of the mirror tube should not extend into the occlusion area of the instrument, so as to avoid injury to the mirror tube. When using this kind of equipment, sometimes doctors extend the endoscope very close to the tissue in order to see clearly the tissue in the occlusal area. When the equipment is occluded, the endoscope does not return and injures the endoscope by mistake. During the operation, if the occlusal mouth of the instrument is all within the scope of observation of the endoscope, such accidents can be avoided.
有些手術窺鏡是在鞘管內使用,在更換其他角度窺鏡或插拔器械時,應注意動作要輕,不可用力過猛。尤其是插拔窺鏡過程中,當遇到阻力拔不動時應仔細查找原因,必要時應連同鞘管一起拔取,不要用蠻力。
Some surgical speculums are used in the sheath. When replacing other angle speculums or inserting and unplugging instruments, it should be noted that the movement should be light and the force should not be too strong. Especially in the process of inserting and pulling out the endoscope, when the resistance is encountered, the reason should be carefully found out. If necessary, the endoscope should be pulled out together with the sheath tube, and do not use brute force.
內窺鏡設備維修
當窺鏡配合激光汽化、高頻電切、微波等光電技術進行手術時,應注意窺鏡前端與點的距離,保證窺鏡前端不被電擊或燒灼。使用這些器械時,主刀醫(yī)生應反復練習,掌握窺鏡圖像中物距和實際物距的關系,確認窺鏡前端與點的近距離,以便在實際手術中應用自如。
When the endoscope is operated with laser vaporization, high-frequency electric cutting, microwave and other photoelectric technologies, attention should be paid to the distance between the front end of the endoscope and the treatment point to ensure that the front end of the endoscope is not shocked or burned. When using these instruments for the first time, the surgeon should practice repeatedly, master the relationship between the object distance and the actual object distance in the endoscope image, and confirm the closest distance between the front end of the endoscope and the treatment point, so as to apply them freely in the actual operation.
目前在耳鼻喉科、骨科的臨床手術中已經(jīng)廣泛使用刨削器來切除病變組織。其刀頭鋒利、硬度高、旋轉速度快、力矩大,如果削到內窺鏡,窺鏡必損無疑。在此類手術中要注意調整沖洗和吸引的速度,隨時保證窺鏡圖像清晰不被血污遮擋,控制刀頭的旋轉部分始終在窺鏡的觀察范圍內,在手術范圍較大時,應先停止刀頭轉動,再移動窺鏡,然后在窺鏡監(jiān)視下移動刀頭,到合適部位后再開機刨削。當感覺到刨刀工作異?;蛘斩韧蝗唤档蜁r,有可能窺鏡已經(jīng)受損,應及時更換,以免造成更大損失。
At present, in the clinical operation of Otolaryngology and orthopedics, planer has been widely used to remove pathological tissue. It has sharp blade, high hardness, fast rotation speed and large torque. If it is cut to the endoscope, the endoscope will be damaged. In this kind of operation, we should pay attention to adjust the speed of flushing and attraction, ensure the clear image of the endoscope at any time and not be covered by blood stain, control the rotating part of the knife head to always be within the scope of observation of the endoscope, when the scope of operation is large, we should stop the rotation of the knife head first, then move the endoscope, then move the knife head under the supervision of the endoscope, and then start the planing after reaching the appropriate position. When it is felt that the planer is working abnormally or the illumination suddenly decreases, it is possible that the endoscope has been damaged, so it should be replaced in time to avoid greater loss.
一般重要手術,應有一套備用窺鏡和關鍵器械,在發(fā)現(xiàn)問題時可以從容地更換;如果使用角度不合適的窺鏡或不配套的器械勉強手術,也容易造成窺鏡的損壞。
For general important operations, there should be a set of spare endoscopes and key equipment, which can be replaced calmly when problems are found; if the use of the endoscope with inappropriate angle or non matching equipment is forced to operate, it is also easy to cause damage to the endoscope.
本文由內窺鏡設備維修為您提供,我們的網(wǎng)站是:http://www.tukouzhao.cn我們將以全心全意的熱情為您提供服務,歡迎您的訪問!
This article is provided by endoscope equipment maintenance. Our website is: http://www.tukouzhao.cn. We will serve you wholeheartedly. Welcome to visit!
s
新聞資訊
聯(lián)系我們
  • 公司公司:匠仁醫(yī)療設備有限公司
  • 匠仁醫(yī)療山東濟南維修部
  • 聯(lián)系人:樊經(jīng)理
  • 電話:13153199508
  • 地址:山東省濟南市槐蔭區(qū)美里東路3000號德邁國際信息產(chǎn)業(yè)園6號樓101-2室
  • 匠仁醫(yī)療湖南長沙維修部
  • 聯(lián)系人:李經(jīng)理
  • 電話:13873135765
  • 地址:湖南省長沙市雨花區(qū)勞動東路820號恒大綠洲14棟2409室
新聞資訊

樊經(jīng)理:13153199508     李經(jīng)理:13873135765

公司地址:山東省濟南市槐蔭區(qū)美里東路3000號德邁國際信息產(chǎn)業(yè)園6號樓101-2室  湖南省長沙市雨花區(qū)勞動東路820號恒大綠洲14棟2409室

     

魯公網(wǎng)安備 37010402441281號