Besides the correct use of medical endoscopes, attention should also be paid to disinfection, cleaning, storage and maintenance. The following endoscopic equipment maintenance personnel mainly talk about the endoscopic cleaning, disinfection, maintenance and other related issues from the environmental requirements for the use of the endoscope, the operation before and after the operation, and the cleaning and disinfection after the operation.
將擦干后的內(nèi)鏡置于多酶洗液中浸泡,時(shí)間按使用說(shuō)明。
Soak the wiped endoscope in multi enzyme lotion for the time specified in the instructions.
使用后立即用流動(dòng)水清洗,并擦干。
Clean with flowing water and wipe dry immediately after use.
器械的軸節(jié)部、彎曲部、管腔內(nèi)用軟毛刷刷洗,刷洗時(shí)注意避免劃傷鏡面。
The shaft joint, bending part and lumen of the instrument shall be brushed with a soft bristle brush, and care shall be taken to avoid scratching the mirror surface.
清洗內(nèi)鏡各部件,管腔應(yīng)當(dāng)用高壓水槍沖洗,可拆卸部分須拆開(kāi)清洗,并用超聲清洗器清洗5~10分鐘。
Clean all parts of the endoscope. The lumen should be flushed with a high-pressure water gun. The detachable parts should be disassembled for cleaning and cleaned with an ultrasonic cleaner for 5-10 minutes.
光纖內(nèi)窺鏡的消毒滅菌方法:消毒應(yīng)用WAYWIN2000醫(yī)用滅菌器滅菌,時(shí)間為30min;或用環(huán)氧乙烷消毒、滅菌。
Disinfection and sterilization method of optical fiber endoscope: use WAYWIN2000 medical sterilizer to sterilize for 30min; Or sterilize with ethylene oxide.
近年來(lái)更多的醫(yī)療機(jī)構(gòu)選用低溫甲醛蒸汽滅菌器,特點(diǎn)是滅菌可靠,對(duì)鏡頭不鈍化影響。適于壓力蒸汽滅菌的內(nèi)鏡或者內(nèi)鏡部件應(yīng)當(dāng)采用壓力蒸汽滅菌,注意按內(nèi)鏡說(shuō)明書(shū)要求選擇溫度和時(shí)間。
In recent years, more medical institutions choose low-temperature formaldehyde steam sterilizer, which is characterized by reliable sterilization and non passivation effect on the lens. Endoscopes or endoscopic components suitable for pressure steam sterilization shall be sterilized by pressure steam, and the temperature and time shall be selected according to the endoscopic instructions.
不能采用壓力蒸汽滅菌的內(nèi)鏡及附件可以使用2%堿性戊二醛浸泡10小時(shí)滅菌。
Endoscopes and accessories that cannot be sterilized by pressure steam can be sterilized by soaking them in 2% alkaline glutaraldehyde for 10 hours.
用消毒液進(jìn)行消毒、滅菌時(shí),有軸節(jié)的器械應(yīng)當(dāng)充分打開(kāi)軸節(jié),帶管腔的器械腔內(nèi)應(yīng)充分注入消毒液。
When disinfecting and sterilizing with disinfectant, the instrument with shaft joint shall fully open the shaft joint, and the instrument cavity with tube cavity shall be fully filled with disinfectant.
在使用醫(yī)療光學(xué)內(nèi)窺鏡之前,一定要認(rèn)真閱讀相關(guān)使用手冊(cè),在了解內(nèi)窺鏡的設(shè)計(jì)、制作工藝、性能等方面的情況的同時(shí),也要注意相關(guān)使用操作說(shuō)明和警告提示文字,防止問(wèn)題的存在,更多相關(guān)內(nèi)容的了解就來(lái)我們網(wǎng)站
http://tukouzhao.cn咨詢吧!
Before using the medical optical endoscope, you must carefully read the relevant user manual. While understanding the design, manufacturing process, performance and other aspects of the endoscope, you should also pay attention to the relevant operating instructions and warning prompts to prevent problems. For more information, please come to our website http://tukouzhao.cn Consult!