內(nèi)窺鏡使用得當,才能發(fā)揮其最大效益。
Only when the endoscope is used properly can its maximum benefit be brought into play.
在日常使用中,如果方法不當,都會出現(xiàn)內(nèi)窺鏡橡皮破損現(xiàn)象,小則影響鏡子正常操作,大則導(dǎo)致鏡子CCD損壞,給醫(yī)院帶來極大的經(jīng)濟損失。
In daily use, if the method is improper, endoscope rubber will be damaged, small will affect the normal operation of the mirror, large will lead to mirror CCD damage, bring great economic losses to the hospital.
現(xiàn)簡單闡述內(nèi)鏡橡皮出現(xiàn)漏水后,如何使用正確方法更換橡皮:
This paper briefly describes how to use the correct method to replace the endoscope rubber after water leakage occurs
1. 用刀片割掉舊橡皮,清潔部位;
Clean the old part with rubber;
2. 在彎曲管兩端涂上一圈環(huán)氧樹脂膠(防止橡皮變形);
2. Apply a circle of epoxy resin adhesive on both ends of the bending pipe (to prevent rubber deformation);
3. 將新橡皮套進專用套筒上,兩頭把橡皮反在套筒上,橡皮里涂上二硫化鉬;
3. Put the new rubber on the special sleeve, turn the rubber on the sleeve at both ends, and coat the rubber with molybdenum disulfide;
4. 用針筒抽真空將橡皮擴張?zhí)走M彎曲管上,松開橡皮兩邊,取走套筒;
4. Use the needle cylinder to vacuum the rubber expansion sleeve into the bending tube, loosen both sides of the rubber and take away the sleeve;
5. 用魚絲扎線,前端是從左到右扎,線寬度為1.5-2MM, 后端分兩道扎,先從右到左扎,再從左到右扎,寬度為 2-2.5MM;用刀片割掉兩邊多余的橡皮;
5. Tie the thread with fish silk. The front end is tied from left to right, with a width of 1.5-2mm. The back end is divided into two ties, first from right to left, and then from left to right, with a width of 2-2.5mm; cut off the excess rubber on both sides with a blade;
6. 用小號狼毫毛筆上膠(AB膠加碳粉);
6. Glue with small Langhao brush (AB glue and carbon powder);
7. 在白熾燈下烘兩小時。
7. Bake in incandescent lamp for two hours.
待膠烘干后,給內(nèi)鏡再進行再次測試,完好后就可以正常使用了。
After the glue is dried, the endoscope is tested again, and it can be used normally.
The above is about endoscope rubber replacement method related content, I hope to help you, want to learn more, welcome to click endoscope equipment repair http://tukouzhao.cn Official website.