內(nèi)窺鏡的幾個特殊故障的預防及檢修

來源:http://www.tukouzhao.cn/ 發(fā)布時間:2023-03-07 瀏覽量:0

由于設備使用的特殊性,內(nèi)窺鏡很容易在使用過程中出現(xiàn)各種故障,例如我們常見的光束被損壞,鉗子管道漏水,水氣管道,噴嘴堵塞,彎曲橡皮漏水等,但是這些故障一般有經(jīng)驗的診療醫(yī)師都是可以自己的,下面內(nèi)窺鏡設備維修小編就將為大家介紹幾種比較特殊的故障情況,希望能有所幫助。
Due to the particularity of the use of the equipment, it is easy for the endoscope to have various faults during the use process, such as the damage of the light beam, the leakage of the pliers pipe, the leakage of the water and gas pipe, the blockage of the nozzle, the leakage of the bending rubber, etc., which are common to us. However, these faults can be solved by the experienced doctors themselves. The following brief introduction will introduce some special fault conditions for you, hoping to help.
故障一
Fault 1
某些型號的內(nèi)窺鏡有時維修后初次使用便會出現(xiàn)無水或是少水的現(xiàn)象。而且在很短的時間內(nèi)視野變得模糊,甚可以看見水,外觀上可以出現(xiàn)彎曲橡皮腫脹。此時用測漏器也無法判斷出什么部位出現(xiàn)問題,打開此鏡會出現(xiàn)鏡體內(nèi)已注滿了水,如果處理不及時,顯微鏡內(nèi)像束大多會產(chǎn)生龜裂,螺線管銅絲出現(xiàn)生銹等情況,嚴重時會造成報廢。
Some models of endoscopes sometimes have no water or little water when they are used for the first time after maintenance. Moreover, in a very short time, the field of vision becomes blurred, even water can be seen, and the appearance can appear bent rubber swelling. At this time, the leak detector can not determine what part has a problem. Opening this mirror will show that the mirror body has been filled with water. If the treatment is not timely, most of the image beams in the microscope will be cracked, and the solenoid copper wire will be rusted. In serious cases, it will cause scrapping.
故障原因
Fault cause
某些內(nèi)窺鏡在操作把下方有一附送水部位,這個部位在修理安裝的過程中,可能會發(fā)生變形或有硬的小雜物,裝配時產(chǎn)生裂痕,水便從此處流入鏡體。所以送氣,送水時會出現(xiàn)水少或無水的情況,而且儲水到一定量時,水壓會使橡皮出現(xiàn)膨脹。
Some endoscopes have an attached water supply part under the operating handle. During the repair and installation process, this part may be deformed or have hard small sundries, which will cause cracks during assembly, and water will flow into the mirror body from this part. Therefore, there will be little or no water when delivering air and water, and when the water is stored to a certain amount, the water pressure will make the rubber expand.
內(nèi)窺鏡設備維修
防止故障方法
Fault prevention method
修理人員在維修完畢后或使用人員在前都應該進行測漏,特別是內(nèi)窺鏡,如果在測漏的過程中發(fā)現(xiàn)噴嘴水汽管出有氣泡冒出而管道完好的情況下,應考慮附送水部位的接口有問題,應重新檢查后安裝,否則待水進入鏡體后側漏器也無法測出。
The repair personnel shall conduct leak detection after the repair or before the user, especially the endoscope. If the nozzle water vapor pipe is found to have bubbles and the pipe is intact during the leak detection, the problem of the interface of the attached water supply part shall be considered first, and it shall be re-inspected before installation, otherwise the leak detector at the back of the mirror body cannot be detected when the water enters.
故障二
Fault 2
機器進水后有時會出現(xiàn)像天空中的彩虹一樣,這種現(xiàn)象也是要在漏水后放置一段時間干燥后才會出現(xiàn),上面已講過導像束是由很多導光束按照一定的順序排列組成的。
The machine will sometimes appear like a rainbow in the sky after water ingress, and this phenomenon will only appear after it is dried for a period of time after water leakage. As mentioned above, the image guide beam is composed of many guide beams arranged in a certain order.
在操作者看起來是一圓形圖像,而實際上每一導光束必須按照正三角形排列才能使像在傳導過程中不致發(fā)生變化,操作者仔細觀察就會知道,特別是經(jīng)過攝像和攝影放大后的圖片會看得更為清楚。我們從整個一條導像束立體去研究,就成了無數(shù)三棱鏡,自然光經(jīng)過三棱鏡后就分為各種不同顏色的單色光,纖維內(nèi)窺鏡出現(xiàn)彩虹的原理和三棱鏡的分光是一樣。
The operator looks like a circular image, but in fact, each guiding beam must be arranged according to an equilateral triangle so that the image will not change during the transmission process. The operator will know after careful observation, especially after the image is taken and magnified, it will be more clear. We study from the whole guiding image beam, and it becomes numerous prism. After passing through the prism, natural light is divided into various monochromatic light of different colors. The principle of rainbow in fiber endoscope is the same as that of prism.
引起設備故障的原因有很多種,這需要醫(yī)師在使用設備中嚴格按照操作規(guī)范去要求自己,這樣才能盡可能地減少設備故障的次數(shù)。更多相關內(nèi)容就來我們網(wǎng)站http://www.tukouzhao.cn留言咨詢,或者直接撥打我們的服務熱線咨詢了解!
There are many reasons for equipment failure, which requires doctors to strictly follow the operating specifications when using the equipment, so as to minimize the number of equipment failures. Come to our website for more relevant content http://www.tukouzhao.cn Leave a message or call our service hotline directly!
s
新聞資訊
聯(lián)系我們
  • 公司公司:匠仁醫(yī)療設備有限公司
  • 匠仁醫(yī)療山東濟南維修部
  • 聯(lián)系人:樊經(jīng)理
  • 電話:13153199508
  • 地址:山東省濟南市槐蔭區(qū)美里東路3000號德邁國際信息產(chǎn)業(yè)園6號樓101-2室
  • 匠仁醫(yī)療湖南長沙維修部
  • 聯(lián)系人:李經(jīng)理
  • 電話:13873135765
  • 地址:湖南省長沙市雨花區(qū)勞動東路820號恒大綠洲14棟2409室
新聞資訊

樊經(jīng)理:13153199508     李經(jīng)理:13873135765

公司地址:山東省濟南市槐蔭區(qū)美里東路3000號德邁國際信息產(chǎn)業(yè)園6號樓101-2室  湖南省長沙市雨花區(qū)勞動東路820號恒大綠洲14棟2409室

     

魯公網(wǎng)安備 37010402441281號