歡迎來到匠仁醫(yī)療設(shè)備有限公司網(wǎng)站!

更多 > 新聞資訊
公司新聞
行業(yè)知識
檢測答疑
圖標(biāo) 首頁 >> 新聞資訊 > 檢測答疑

檢測答疑

消化內(nèi)窺鏡出現(xiàn)“吸引”故障什么原因

來源:http://www.tukouzhao.cn/ 發(fā)布時間:2023-02-11 瀏覽量:0

如今消化內(nèi)鏡已廣泛應(yīng)用于各級醫(yī)院,為使患者享受高質(zhì)量的醫(yī)療服務(wù),同時降低科室昂貴的維修費(fèi)用,內(nèi)鏡故障的及時處理及日常維護(hù)保養(yǎng)顯得尤為必要。那么消化內(nèi)窺鏡出現(xiàn)“吸引”故障什么原因?內(nèi)窺鏡設(shè)備維修人員為您解答。
Nowadays, digestive endoscopy has been widely used in hospitals at all levels. In order to enable patients to enjoy high-quality medical services and reduce the expensive maintenance costs of departments, timely treatment and daily maintenance of endoscope failures are particularly necessary. So what is the reason for "attraction" failure of digestive endoscope? The endoscope equipment maintenance personnel will answer for you.
常見故障原因:
Common fault causes:
(1)吸引按鈕損壞,上面的O型圈不全導(dǎo)致密封性不好;
(1) The suction button is damaged, and the O-ring on it is incomplete, resulting in poor sealing;
(2)食物殘渣或糞便血塊等異物被吸入插入部的通管,多發(fā)生在管道的狹窄處及轉(zhuǎn)彎處;
(2) Foreign matters such as food residues or fecal blood clots are inhaled into the pipe of the insertion part, which often occurs at the narrow and turning points of the pipe;
內(nèi)窺鏡設(shè)備維修
(3)錯誤抽出清洗刷及附件的手法,磨損吸引安裝座,導(dǎo)致漏氣,吸引不暢。
(3) The wrong way to draw out the cleaning brush and accessories will wear the suction mounting seat, resulting in air leakage and poor suction.
預(yù)防處理措施:使用前及時更換損壞的吸引按鈕,堵塞時可用洗耳球?qū)?zhǔn)吸引按鈕孔用力吹出,或者用清潔刷從鉗子口插入清洗吸引管道,完畢后要在吸引按鈕膠圈上抹上硅油,并重新安裝好。抽出附件與清洗刷時,手指應(yīng)拖住下方再水平抽出,避免與按鈕安裝座發(fā)生摩擦。
Preventive measures: replace the damaged suction button in time before use. If it is blocked, use the ear wash ball to aim at the suction button hole and blow it out forcefully, or use the cleaning brush to insert the cleaning suction pipe from the pliers mouth. After completion, apply silicone oil on the rubber ring of the suction button and reinstall it. When pulling out the accessories and cleaning brush, the fingers should be pulled down and then pulled out horizontally to avoid friction with the button mounting base.
消化內(nèi)窺鏡昂貴,內(nèi)管道較多,不合理的使用及保養(yǎng)易造成內(nèi)窺鏡的損壞,使用及維護(hù)人員須定期進(jìn)行專門的培訓(xùn),避免人為損壞及故障,發(fā)現(xiàn)問題及時處理、送修,避免問題進(jìn)一步惡化。更多相關(guān)事項就來我們網(wǎng)站http://www.tukouzhao.cn咨詢吧!
The digestive endoscope is expensive and has many internal pipes. Improper use and maintenance can easily cause damage to the endoscope. The use and maintenance personnel must regularly carry out special training to avoid man-made damage and failure, and timely handle and send for repair if problems are found to avoid further deterioration of the problem. Come to our website for more relevant matters http://www.tukouzhao.cn Ask!
s
新聞資訊
聯(lián)系我們
  • 公司公司:匠仁醫(yī)療設(shè)備有限公司
  • 匠仁醫(yī)療山東濟(jì)南維修部
  • 聯(lián)系人:樊經(jīng)理
  • 電話:13153199508
  • 地址:山東省濟(jì)南市槐蔭區(qū)美里東路3000號德邁國際信息產(chǎn)業(yè)園6號樓101-2室
  • 匠仁醫(yī)療湖南長沙維修部
  • 聯(lián)系人:李經(jīng)理
  • 電話:13873135765
  • 地址:湖南省長沙市雨花區(qū)勞動?xùn)|路820號恒大綠洲14棟2409室
新聞資訊

樊經(jīng)理:13153199508     李經(jīng)理:13873135765

公司地址:山東省濟(jì)南市槐蔭區(qū)美里東路3000號德邁國際信息產(chǎn)業(yè)園6號樓101-2室  湖南省長沙市雨花區(qū)勞動?xùn)|路820號恒大綠洲14棟2409室

     

魯公網(wǎng)安備 37010402441281號